PHYS-397 · Lesson 5 of 12 · 22 min

Electronics for physicists I: circuits, impedance and measurement instruments

Almost every modern measurement ends as a voltage. A photodiode turns light into current, a thermocouple turns temperature into a few millivolts, a strain gauge turns stretch into a change in resistance, a particle detector turns ionization into a pulse. Between the sensor and the number in your notebook lies electronics, and an experimenter who treats it as a black box will sooner or later be fooled. This lesson reviews the circuit ideas a physicist needs, extends them to alternating current with complex impedance, and studies the instruments of the bench: the multimeter, the oscilloscope and the signal generator. Its main practical lesson is that every instrument loads the circuit it measures. PHYS-131 and PHYS-382 cover the underlying electromagnetism; here we concentrate on measurement.

Circuit fundamentals

Kirchhoff's current law says that the currents entering a node sum to zero, a statement of charge conservation; his voltage law says that the voltages around any closed loop sum to zero, a statement that the electrostatic field is conservative. With Ohm's law, \(V = IR\), these two laws let you analyse any network of resistors and sources. Two results are used constantly. Resistors in series add, \(R = R_1 + R_2\); in parallel their conductances add, \(1/R = 1/R_1 + 1/R_2\). And a voltage divider, two resistors in series across a source \(V_s\), produces at the junction

\[V_{\text{out}} = V_s\,\frac{R_2}{R_1 + R_2}.\]

Thévenin's theorem says that any linear network, seen from two terminals, behaves like a single voltage source \(V_{\text{Th}}\) in series with a single resistance \(R_{\text{Th}}\). For the divider, \(V_{\text{Th}}\) is the open-circuit output and \(R_{\text{Th}} = R_1R_2/(R_1 + R_2)\), the two resistors in parallel. The Thévenin resistance is also called the output impedance of the source. It is the key to understanding loading.

Loading: the instrument changes the measurement

A voltmeter connected across two points draws a little current, and so acts like a resistor \(R_{\text{in}}\), its input impedance, in parallel with whatever it is measuring. The reading is then \(V_{\text{Th}}R_{\text{in}}/(R_{\text{Th}} + R_{\text{in}})\) instead of \(V_{\text{Th}}\). The error is small only if \(R_{\text{in}} \gg R_{\text{Th}}\). A modern digital multimeter has an input resistance of about 10 MΩ on its voltage ranges; an oscilloscope input is usually 1 MΩ in parallel with about 15 pF; an old moving-coil meter rated at 20 kΩ per volt has only 200 kΩ on its 10 V range. An electrometer, built to measure the potentials of high-impedance sources such as ionization chambers or pH electrodes, can exceed \(10^{14}\) Ω.

Left: a circuit diagram of a 10 volt source with two 1 megohm resistors in series; a voltmeter symbol labelled R in is connected across the lower resistor. Right: a bar chart of the reading for four instruments: moving-coil meter about 1.4 volts, oscilloscope about 3.3 volts, digital multimeter about 4.8 volts, electrometer 5.0 volts. A dashed line marks the ideal 5.0 volts.
Measuring the midpoint of a 1 MΩ divider. A meter whose input resistance is not much larger than the source resistance reads low

The same idea applies to current measurements, where an ammeter's own resistance, its burden, adds to the circuit and reduces the current it measures, and to every sensor: a thermometer exchanges heat with what it measures, a probe disturbs the field it samples. The general rule is to make the measuring device's influence small compared with the quantity measured, or to correct for it. The program below quantifies loading for a divider of two 1 MΩ resistors, whose true midpoint voltage is 5.000 V.

Alternating current and complex impedance

For sinusoidal signals, write voltages and currents as the real parts of complex exponentials, \(V(t) = \mathrm{Re}(\tilde{V}e^{i\omega t})\). Then resistors, capacitors and inductors all obey a generalized Ohm's law \(\tilde{V} = Z\tilde{I}\), with impedances

\[Z_R = R, \qquad Z_C = \frac{1}{i\omega C}, \qquad Z_L = i\omega L.\]

These follow from \(V = IR\), \(I = C\,dV/dt\) and \(V = L\,dI/dt\): differentiating \(e^{i\omega t}\) multiplies it by \(i\omega\). Impedances combine in series and parallel exactly like resistances, so the whole machinery of DC circuits carries over, with complex numbers. The magnitude \(|Z|\) gives the ratio of voltage and current amplitudes, and the phase of \(Z\) gives how far the voltage leads the current. A capacitor's impedance falls with frequency, so it blocks DC and passes high frequencies; an inductor does the opposite. At 1 kHz, a 100 nF capacitor has an impedance magnitude of 1.59 kΩ and a 100 mH inductor 628 Ω. In the program, a 1 kΩ resistor in series with the capacitor has \(|Z| = 1.88\) kΩ with the voltage lagging the current by 58 degrees.

The Wheatstone bridge

Many sensors produce a small change in a resistance: a platinum thermometer changes by 0.39 Ω per kelvin, a metal-foil strain gauge by about twice the strain times its resistance. Measuring a 0.1 parts in 1000 change in a resistance directly would require a meter of extraordinary accuracy. The Wheatstone bridge, popularized by Charles Wheatstone in 1843, avoids the problem by measuring a difference. Four resistors form two dividers fed from the same source; the output is the voltage between their midpoints. When the bridge is balanced, the output is zero whatever the supply voltage, and a small change \(\delta R\) in one arm produces an output

\[V_{\text{out}} \approx V_b\,\frac{\delta R}{4R},\]

which can be amplified without the large steady voltage that would otherwise swamp it. The bridge is an example of a null method, comparing an unknown with a standard by adjusting for zero output. Null methods are powerful because the detector only needs to be sensitive, not calibrated.

Loading a voltage divider with different instruments, complex impedances at 1 kHz, and the output of a strain-gauge bridge
import math, cmath
# 1. A voltage divider measured by meters of different input resistance.
Vs, R1, R2 = 10.0, 1.0e6, 1.0e6
ideal = Vs*R2/(R1 + R2)
for name, Rin in (("moving-coil meter, 20 kohm/V, 10 V range", 2.0e5), ("oscilloscope, 1 Mohm", 1.0e6), ("digital multimeter, 10 Mohm", 1.0e7), ("electrometer, 1e14 ohm", 1.0e14)):
    Rp = R2*Rin/(R2 + Rin)
    V = Vs*Rp/(R1 + Rp)
    print(f"{name:44s}: reads {V:6.3f} V (ideal {ideal:.3f} V, error {round((V/ideal - 1)*100, 1) + 0.0:+6.1f} parts per hundred)")
# 2. Complex impedances at 1 kHz.
f = 1000.0; w = 2*math.pi*f
R, C, Lh = 1000.0, 100e-9, 0.10
for name, Z in (("R + C in series", R + 1/(1j*w*C)), ("R + L in series", R + 1j*w*Lh), ("R parallel C", 1/(1/R + 1j*w*C))):
    print(f"{name:16s}: |Z| = {abs(Z):8.1f} ohm, phase = {math.degrees(cmath.phase(Z)):6.1f} degrees")
# 3. Wheatstone bridge: output for a small change in one arm (strain gauge, 120 ohm, gauge factor 2.0).
Vb, Rg = 5.0, 120.0
for strain in (1e-5, 1e-4, 1e-3):
    dR = Rg*2.0*strain
    Vout = Vb*((Rg + dR)/(2*Rg + dR) - 0.5)
    print(f"strain {strain:.0e}: bridge output {Vout*1e6:9.2f} microvolts (linear estimate {Vb*dR/(4*Rg)*1e6:9.2f})")
Output
moving-coil meter, 20 kohm/V, 10 V range    : reads  1.429 V (ideal 5.000 V, error  -71.4 parts per hundred)
oscilloscope, 1 Mohm                        : reads  3.333 V (ideal 5.000 V, error  -33.3 parts per hundred)
digital multimeter, 10 Mohm                 : reads  4.762 V (ideal 5.000 V, error   -4.8 parts per hundred)
electrometer, 1e14 ohm                      : reads  5.000 V (ideal 5.000 V, error   +0.0 parts per hundred)
R + C in series : |Z| =   1879.6 ohm, phase =  -57.9 degrees
R + L in series : |Z| =   1181.0 ohm, phase =   32.1 degrees
R parallel C    : |Z| =    846.7 ohm, phase =  -32.1 degrees
strain 1e-05: bridge output     25.00 microvolts (linear estimate     25.00)
strain 1e-04: bridge output    249.98 microvolts (linear estimate    250.00)
strain 1e-03: bridge output   2497.50 microvolts (linear estimate   2500.00)

The moving-coil meter reads 1.43 V instead of 5.00 V, an error of 71 parts in 100, because its input resistance is comparable to the source resistance. The oscilloscope, at 1 MΩ, reads exactly a third low. Even the 10 MΩ multimeter is 5 parts in 100 low, which would be a serious systematic error in a precise experiment. Only the electrometer reads the true value. The bridge results show the linear approximation \(V_b\,\delta R/4R\) is excellent for small strains and still within 1 part in 1000 at a strain of \(10^{-3}\): a strain of \(10^{-5}\), a change in length of 10 μm per metre, produces 25 μV, easily measured after amplification.

Instruments of the bench

The digital multimeter measures DC and AC voltage, current and resistance. Its specifications deserve reading: accuracy is usually quoted as a fraction of reading plus a number of counts in the last digit, and AC ranges are accurate only within a stated frequency band. A true-rms meter computes the root mean square of any waveform; a cheaper average-responding meter assumes a sine wave and reads wrongly for other shapes. For resistance, a four-wire measurement passes current through one pair of leads and measures voltage with another, so that the resistance of the leads and contacts drops out; this is essential below about 10 Ω.

The oscilloscope displays voltage against time. Its essential specifications are bandwidth, the frequency at which a sine wave is displayed 3 dB low, and sampling rate. A 10:1 passive probe raises the input impedance to 10 MΩ and reduces capacitive loading, at the price of a tenfold attenuation; it must be compensated, by adjusting a small capacitor until a square calibration wave looks square, or it will distort fast edges. The trigger, which starts each sweep at the same point on a repeating signal, is the control most worth mastering. Signal generators supply sine, square and pulse waveforms and are the standard way to test a circuit's response; their output impedance, usually 50 Ω, forms a divider with the load like any source.

Calibration and traceability

An instrument's reading is only as good as its calibration. Every reputable meter is calibrated against a more accurate standard, which was calibrated against a better one, and so on up a chain that ends at a national metrology institute, the National Research Council of Canada, which realizes the SI units from their defining constants. This unbroken chain of comparisons, each with a stated uncertainty, is called traceability. For voltage, the top of the chain is the Josephson voltage standard, an array of superconducting junctions that produces voltages fixed by frequency and the ratio \(h/2e\); for resistance it is the quantum Hall effect, which gives resistances fixed by \(h/e^2\). Calibration drifts with time, temperature and use, so laboratory instruments carry calibration certificates with dates and are recalibrated on a schedule. In a research experiment, it is good practice to check an instrument against a known reference before and after a critical run: a standard resistor, a voltage reference chip, a known signal from a generator. If the check fails, you know the run is suspect; if you never checked, you will not know. The same logic applies to every sensor in the laboratory. A thermocouple should be checked at the ice point, a photodiode against a calibrated power meter, a balance with reference masses.

Grounding and interference

Many puzzling signals in the laboratory are interference. A loop of wire picks up the 60 Hz magnetic field of the building wiring, the frequency of the North American power grid, and its harmonics. Two instruments connected to ground at different points form a ground loop, in which a small current flows and appears as an offset or hum. Long cables act as antennas for radio-frequency noise. The remedies are standard: keep loops small, twist pairs of signal wires, use shielded coaxial cable with the shield grounded at one end, connect all grounds at a single point, and keep sensitive inputs away from transformers and motors. Lesson 6 shows how to reduce noise further by amplification and filtering, and lesson 7 by lock-in detection.

Worked problem

Correcting for meter loading

A 10 MΩ digital meter reads 4.762 V at the midpoint of a divider made of two 1.00 MΩ resistors across 10.000 V. Show that the reading is consistent with loading and recover the true open-circuit voltage.

Show solution
  1. Thévenin equivalent at the midpoint: \(V_{\text{Th}} = 5.000\) V and \(R_{\text{Th}} = 0.500\) MΩ.
  2. Predicted reading: \(5.000 \times 10/(10 + 0.5) = 4.762\) V, matching the measurement.
  3. Inverting: \(V_{\text{Th}} = V_{\text{read}}(R_{\text{Th}} + R_{\text{in}})/R_{\text{in}} = 4.762 \times 1.05 = 5.000\) V.
  4. The correction requires knowing \(R_{\text{in}}\); check it with the meter's specification or by a second measurement with a different known load.

Answer: The 0.238 V shortfall is exactly the loading error; correcting gives 5.000 V. Better still, use an instrument with input impedance much larger than 0.5 MΩ.

Common mistakes

  • MistakeAssuming a voltmeter draws no current.

    InsteadEvery meter has a finite input impedance; compare it with the source's Thévenin resistance.

  • MistakeMeasuring small resistances with two wires.

    InsteadLead and contact resistances add; use a four-wire measurement.

  • MistakeUsing an uncompensated oscilloscope probe.

    InsteadAdjust it on the calibration square wave before trusting fast signals.

  • MistakeForgetting the phase in AC circuits.

    InsteadCombine impedances as complex numbers; magnitudes alone do not add.

  • MistakeGrounding a shield at both ends.

    InsteadThat can create a ground loop; ground at one point unless high frequencies demand otherwise.

Summary and key terms

  • Kirchhoff's laws and Ohm's law analyse any resistor network; Thévenin's theorem reduces it to a source and a resistance.
  • An instrument with input impedance \(R_{\text{in}}\) reads \(V_{\text{Th}}R_{\text{in}}/(R_{\text{Th}} + R_{\text{in}})\).
  • Impedances \(R\), \(1/i\omega C\) and \(i\omega L\) combine like resistances in AC circuits.
  • A Wheatstone bridge measures small resistance changes as a difference, \(V_b\,\delta R/4R\).
  • Read instrument specifications: input impedance, accuracy, bandwidth and true-rms response.
  • Small loops, twisted pairs, shielding and single-point grounding reduce interference.
Thévenin equivalent
A single source and series resistance that reproduce a linear network at two terminals.
input impedance
The effective impedance an instrument presents to the circuit it measures.
loading
The change in a circuit caused by connecting a measuring instrument.
impedance
The complex ratio of voltage to current for a sinusoidal signal.
Wheatstone bridge
Four resistors arranged as two dividers whose output difference measures a small change.
ground loop
A closed path through two ground connections in which unwanted current flows.

Practice problems

Work each problem on paper first, then reveal the answer to check.

  1. Find the Thévenin resistance at the midpoint of a divider made of 4.7 kΩ and 10 kΩ.

    Show answer

    \(4.7 \times 10/14.7 = 3.2\) kΩ.

  2. A source has a Thévenin resistance of 100 kΩ. What input impedance is needed for a loading error below 1 part in 1000?

    Show answer

    \(R_{\text{in}} > 1000 \times 100\) kΩ, about 100 MΩ.

  3. Find the impedance of a 10 μF capacitor at 50 Hz and at 50 kHz.

    Show answer

    \(1/(2\pi fC)\): 318 Ω and 0.32 Ω.

  4. A Wheatstone bridge with 350 Ω arms and a 10 V supply has one arm change by 0.70 Ω. Estimate the output.

    Show answer

    \(10 \times 0.70/(4 \times 350) = 5.0\) mV.

  5. Why does a 10:1 oscilloscope probe reduce loading?

    Show answer

    It raises the input resistance to 10 MΩ and lowers the capacitance seen by the circuit.

  6. A loop of area 0.01 m\(^2\) sits in a 60 Hz field of amplitude 1 μT. Find the amplitude of the induced voltage.

    Show answer

    \(\omega BA = 377 \times 10^{-6} \times 0.01 = 3.8\) μV.

Lesson quiz · 5 questions

Score 60% or more to complete this lesson.

Take the quiz